LMC-10
Contact luminance meter
The LMC-10 contact luminance meter is designed for the luminance measurement of the self-luminescent surfaces such as matrices of monitors and TV sets, negatoscopes, reading panels, etc.
Convenience
The meter can be attached to the P-200 control panel or to a PC through the RF-200C interface. The LMC-10 contact luminance meter is equipped with the multifunctional key which is used to trigger a measurement, save results or operate the chosen measurement procedure. Owing to the in-built application, which is used to test grayscale mapping curve according to DICOM GSDF function, the meter is indispensable for testing medical imaging monitors and negatoscopes.
Precision
It ensures the accurate measurement, regardless of the spatial distribution of the measured luminance and the backlight system of the surface being tested. Combined with the advanced algorithms of the signal processing, its optical and measurement systems ensure high accuracy and repeatability of measurements.
Technical data
- Spectral range:
- V (λ) CIE
- Measurement ranges:
- 50 000 cd / m2
500 cd / m2 - Accuracy class:
- A (CIE, DIN 5032-7)
- Field aperture:
- 1º
- Measurement field:
- Ø10 mm
- Spectral matching f1':
- ≤ 3%
- Directional matching f2g:
- ≤ 3%
- Effect of the surrounding field f2u:
- ≤ 1.5% (CIE)
- Measurement resolution:
- 0.01 cd / m2
- Temperature range:
- -10 ÷ + 50ºC
- Relative humidity:
- ≤ 90%
- Weight:
- 150 g